Detail of Recall

Type of Product:Device
Medsafe Reference:19002
Brand Name:Philips Brilliance 64 and Ingenuity CT Scanners
Model:Brilliance 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128
Affected:Versions 4.0.x.xxx to 4.1.4
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Some scans unexpectedly set the Axial (Z axis) scan length to a greater length than the user set in the plan. There is a potential for undesired radiation to areas not intended to be scanned.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/8/2015