Detail of Recall

Type of Product:Device
Medsafe Reference:18972
Brand Name:Philips Trauma Diagnost, C-arm
Model:All Trauma Diagnost systems delivered prior to 8th of July 2007 are affected.
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:If two of the three fixation bolts of the Trauma supporting block connecting the Trauma arm with the CS telescope arm are cracked, the Trauma arm is no longer fastened securely. Subsequently, the 3rd bolt may easily break during normal use and the Trauma arm would fall down.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/8/2015