Detail of Recall

Type of Product:Device
Medsafe Reference:18964
Brand Name:Biocartis NVIdylla™ KRAS Mutation Test
Model:A0020
Affected:2207
Software version:
Recalling Organisation:Abacus ALS Limited
55b Richard Pearse Drive
Airport Oaks
Auckland 2022
Contact Information:0800 222 170
Manufacturer:Biocartis NV
Issue:Two false negative results were obtained during initial laboratory validation testing in a Belgian clinical laboratory. A "No mutation detected in KRAS codon 12,13,59,61,117,146'" result was reported on a reference material of Horizon, containing 5 % allelic frequency of G12A, using the Idylla™ KRAS Mutation Test, lot 2207.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/8/2015