Detail of Recall

Type of Product:Device
Medsafe Reference:18963
Brand Name:Leica ASP 200S Analyser
Model:14048043625
Affected:S/N 5457
Software version:
Recalling Organisation:Leica Microsystems Pty Ltd, New Zealand Branch
3/33 Spartan Road
Takanini
Auckland 2105
Contact Information:021 0841 7323
Manufacturer:Leica Instruments Singapore Pte Ltd
Brand Name:Leica ASP 300S Analyser
Model:14047643514
Affected:S/Ns 5832, 5834, 5853
Software version:
Recalling Organisation:Leica Microsystems Pty Ltd, New Zealand Branch
3/33 Spartan Road
Takanini
Auckland 2105
Contact Information:021 0841 7323
Manufacturer:Leica Instruments Singapore Pte Ltd
Issue:The manufacturer recently became aware that these devices have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm not to function correctly in the case of an instrument failure.
If there is an instrument failure during processing and the remote alarm is relied upon then the missing alarm signal could lead to suboptimal quality tissue processing. In some cases, a re-biopsy of the patient could be necessary in order to obtain another viable tissue sample.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/8/2015