Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18963 |
| Brand Name: | Leica ASP 200S Analyser |
| Model: | 14048043625 |
| Affected: | S/N 5457 |
| Software version: | |
| Recalling Organisation: | Leica Microsystems Pty Ltd, New Zealand Branch 3/33 Spartan Road Takanini Auckland 2105 |
| Contact Information: | 021 0841 7323 |
| Manufacturer: | Leica Instruments Singapore Pte Ltd |
| Brand Name: | Leica ASP 300S Analyser |
| Model: | 14047643514 |
| Affected: | S/Ns 5832, 5834, 5853 |
| Software version: | |
| Recalling Organisation: | Leica Microsystems Pty Ltd, New Zealand Branch 3/33 Spartan Road Takanini Auckland 2105 |
| Contact Information: | 021 0841 7323 |
| Manufacturer: | Leica Instruments Singapore Pte Ltd |
| Issue: | The manufacturer recently became aware that these devices have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm not to function correctly in the case of an instrument failure. If there is an instrument failure during processing and the remote alarm is relied upon then the missing alarm signal could lead to suboptimal quality tissue processing. In some cases, a re-biopsy of the patient could be necessary in order to obtain another viable tissue sample. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/8/2015 |