Detail of Recall

Type of Product:Device
Medsafe Reference:18956
Brand Name:Medcomp Hemo-cath®LT and Hemo-cath®ST Hemodialysis Catheters
Model:LT and ST
Affected:ECS1316 ECS1320 and ECS1324
Software version:
Recalling Organisation:AA-Med NZ Ltd
Level 7, 54 Gill Street
New Plymouth 4310
Contact Information:
Manufacturer:Medcomp Inc
Issue:Manufacturer notification of potential safety issue affecting Medcomp Hemo-Cath LT and Hemo-Cath ST Hemodialysis catheters which may develop holes during use. Investiagtion determined that during the insertion procedure engaging the extension line clamp over the guidewire may damage the inside of the tubing. A crack in the tubing on the inside can propagate over time and eventually travel through thickness of the tubing developong into a hole that leaks.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/8/2015