Detail of Recall

Type of Product:Device
Medsafe Reference:18951
Brand Name:Siemens Advia Centaur XPT
Model:10711433
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has identified multiple issues with the ADVIA Centaur® XPT system software V1.0.1 (Bundle 1.0.912 SMN 10819704) and V1.0.2 (Bundle 1.0.1086 SMN 11219806).
When running ThCG samples with an initial dilution of 1 to 200, without processing neat specimens, it is potentially possible to miss a positive ThCG value where the neat value is less than 1000. A lookback of 48 hours is recommended for negative ThCG samples if neat values were not initially run for ThCG and large dilutions were employed.
There is the extremely unlikely potenial to transpose patient results when running in rack order mode without positive sample identificiation.
Similarly a lookback of 48 hours for any instruments running in this mode is recommended.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/8/2015