Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18950 |
| Brand Name: | GE Lightspeed CT Scanners with MDAS 16 5V power supply |
| Model: | LightSpeed Ultra, LightSpeed Plus, LightSpeed QX/i, LightSpeed 16 |
| Affected: | Manufactured from 2003-2006 |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | GE Healthcare has become aware that the Service Procedure used to replace a power supply on the rotating side of the gantry specifies the wrong torque for the mounting hardware. Because of this error, the three mounting bolts securing the power supply could have been over-torqued when the component was replaced since installation. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were to be expelled. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/8/2015 |