Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18948 |
| Brand Name: | Siemens RapidPoint 500 System |
| Model: | 10492730, 10696855, 10696857, 10697306 |
| Affected: | All units |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd 55 Hugo Johnston Drive Penrose Auckland |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port. This could cause the message received by the LIS (data management system) to include either * Duplicate data * Missing data * Data from a different patient. Siemens has not observed this scenario and the expected frequency of occurrence is extremely unlikely. If this issue occurs, only one analyte has the potential to be affected (i.e. if a panel of analytes is ordered, only one could be affected). |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/8/2015 |