Detail of Recall

Type of Product:Device
Medsafe Reference:18923
Brand Name:Siemens N Latex Beta2-Microglobulin & DV Flex Reagent B2MIC
Model:OQWU155 & OQWU215
Affected:44696, 44800, 44834; and 44697
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Products GmbH
Issue:Siemens Healthcare Diagnostics has confirmed complaints about calibration or QC validation issues as well as 'Abnormal Assay Error' flags when using the above mentioned lots. Affected reagent vials and reagent flexes show partial flocculation and share the same production bulk.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/7/2015