Detail of Recall

Type of Product:Device
Medsafe Reference:18922
Brand Name:Siemens Dimension & Dimension Vista Creatinine
Model:DF33A & K1033
Affected:All in-date and future lots are affected
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has determined that Dimension Vista® Creatinine (CREA) may produce depressed results up to 22% (or approximately 27 µmol/L) for a creatinine result at 133 µmol/L when hemolysis exceeds 200 mg/dL [0.12 mmol/L] of haemoglobin. The percent haemoglobin interference with Dimension Vista® CREA is dependent on the creatinine concentration; at a creatinine concentration of 442 µmol/L, haemoglobin interference at concentrations up to a 1000 mg/dL [0.62 mmol/L] is < 10%. Alternately, with an approximate bias of 27 µmol/L, the percent haemoglobin interference may be higher at low creatinine concentrations (< 71 µmol/L).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/7/2015