Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18919 |
| Brand Name: | Siemens Artis Systems with Artis table |
| Model: | 5767004, 5767012, 5767020, 5767053, 10759275, 10759276, 10759277, 10759278 |
| Affected: | Multiple serial number ranges |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | If the tabletop is fully extended at the head end and there is a large amount of liquid on the table surface, the liquid can make its way from the foot end of the tabletop onto the top cover of the table column where it can seep into the table through a gap. If this causes electronic components to become contaminated, table movements may no longer be possible. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/7/2015 |