Detail of Recall

Type of Product:Device
Medsafe Reference:18919
Brand Name:Siemens Artis Systems with Artis table
Model:5767004, 5767012, 5767020, 5767053, 10759275, 10759276, 10759277, 10759278
Affected:Multiple serial number ranges
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:If the tabletop is fully extended at the head end and there is a large amount of liquid on the table surface, the liquid can make its way from the foot end of the tabletop onto the top cover of the table column where it can seep into the table through a gap. If this causes electronic components to become contaminated, table movements may no longer be possible.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/7/2015