Detail of Recall

Type of Product:Device
Medsafe Reference:18899
Brand Name:Iris iChem Velocity Automated Urine Chemistry System
Model:800-3042, 800-3043, 800-3044, 800-3075, 800-3076, 800-3564, 800-7100, 800-7101, 800-7102, 800-7103, 800-7106, 800-7157, 800-7162, 800-7163, 800-7166
Affected:Manufactured between Apr-2011 and Jun-2015
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Iris International
Issue:IRIS has confirmed the iChem Velocity Automated Urine Chemistry Systems do not maintain the on-board strip stability claim of 10 days at 18-28 degrees C at 20-80% relative humidity as stated in the product labeling. The on-board strip stability is affected by temperature and humidity in the laboratory. If a laboratory operates outside of the defined parameters , test results may be affected as indicated below:
* False low or false negative results: Ketone, Blood
* False high or false positive results: Bilirubin, Urobilinogen, Glucose, Nitrite
* False high results: Leukocytes
Ascorbic Acid, Protein and pH are not affected.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2015