Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18889
Brand Name:Siemens E.cam System
Model:
Affected:Installed prior to 31-Dec-2004
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer has been advised of a report of unintended detector motion caused by the simultaneous failure of two components. This unique combination of failures caused the axis brake to be ineffective, leading to the detector motion. Siemens will install an improved brake control circuit to correct this condition.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/7/2015