Detail of Recall

Type of Product:Device
Medsafe Reference:18872
Brand Name:Q Core Sapphire Infusion Pump
Model:Sapphire
Affected:All units
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Q Core Medical Ltd
Issue:The manufacturer has received complaints from customers that when using the Epidural mode with patient bolus, the amount of medication delivered may be greater than programmed. Q Core Medical is adding to the User Manual the parameters of the Reference Catheter to allow precise correlation between the catheter dimensions, the back pressure and the deviation in flow that the catheter may cause.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/7/2015