Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18860
Brand Name:Tornier Aequalis IM Nail Targeting Jig
Model:9020060
Affected:Lots: All lots - (DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102)
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Tornier SA
Issue:The targeting jig mast is separating from the jig boom. Any separation will impact alignment and can affect the ability to fixate the screws.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Sponsor Control
Date Commenced:14/7/2015
Hide menus