Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18837
Brand Name:Synthes Reamer / Irrigation / Aspiration tube assemblies
Model:314.745S, 314.746S
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Synthes GmbH
Brand Name:Synthes Reamer
Model:314.742
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Synthes GmbH
Issue:It was reported that the drive shaft, tube assembly, and reamer head have the potential to break when inccorrectly assembled or used improperly. Precautionary statements are being added to the "Reamer / Irrigator / Aspirator (RIA) Surgical Technique Guide".
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/7/2015
Hide menus