Detail of Recall

Type of Product:Device
Medsafe Reference:18834
Brand Name:Siemens Advia Centaur XPT System
Model:10711433
Affected:All serial numbers manufactured from Oct-2014 to Feb-2015
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens has identified a number of workflow issues with Advia Centaur XPT system software V1.0.1 and V1.0.2:
* The system does not display any indication of a missing required Probe Wash pack on the results screen.
* The data shown on the "Utilities->Assay Utilization" screen is incorrect for the totals of calibrators run on the instrument
* The data shown on the "Utilities->Assay Utilization" screen and the printout report for Assay Utiliization may not match
The following issues are related to instruments connected to a Lab Automation System (LAS):
* If the instrument receives work orders while transition from the In-Process state to the Ready state, the sample will be routed to the instrument but may not be processed
*The system may stop processing samples from the LAS and hold a tube at the LAS sampling station
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/7/2015