Detail of Recall

Type of Product:Device
Medsafe Reference:18832
Brand Name:Siemens Advia Centaur XPT System
Model:10711433
Affected:All serial numbers manufactured from Oct-2014 to Feb-2015
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has identified the following issue on the ADVIA Centaur® XPT system :
· The time of initiation for onboard stability (OBS) on the ADVIA Centaur® XPT system is calculated when probe wash reagent packs are pierced (used by the system), not when loaded onto the instrument, for probe wash reagents PW2, PW3 and PW4.
· Because onboard stability (OBS) is calculated when the probe wash reagent pack is pierced, not when loaded onto the instrument, the probe wash reagent pack may be on the instrument longer than described in the Instructions For Use (IFU).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/7/2015