Type of Product:Device
Medsafe Reference:18830
Brand Name:Medline Custom Procedure Kit, incorporating Unomedical suction tubing
Model:SOTNZAH04B
Affected:CS1405587, CS1417569, CS1418438, CS1418710, CS1419278, CS1420864, CS1422092, CS1423500, CS1423501, CS1423955, CS1527665, CS1528112, CS1528978, CS1529680, CS1529789, CS1531866
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Medline Assembly Australia Pty Ltd
Issue:Medline is recalling specified lots of the Unomedical Suction tubing. This tubing is a component in the Medline procedure pack product code SOTNZAH04B.
An internal assessment of ConvaTec's product complaints confirmed that these devices were not meeting customer expectations. The connector parts for application to the suction device in these lots have failed to meet their required reliability. The connector parts of the specified lots have a higher probability to crack once applied to suction device outlets. Cracking may result in loss of suction during an operation. This could result in reduced surgical site visibility (due to poor removal of fluid) and may make stemming of bleeding more difficult.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2015