Detail of Recall

Type of Product:Device
Medsafe Reference:18826
Brand Name:Draeger Babylog VN500 Ventilator with optional PS500 installed
Model:VN500
Affected:All units with optional PS500 installed
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Brand Name:Draeger Infinity V500 Ventiilator with optional PS500 installed
Model:V500
Affected:All units with optional PS500 installed
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Brand Name:Draeger Evita V500 Ventilator, with optional PS500 installed
Model:V500
Affected:All units with optional PS500 installed
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:Draeger has become aware of an event in which the battery capacity of the optional PS500 of an IACS Workstation Critical Care (Evita Infinity V500, Evita V300 and Babylog VN500 ventilators) did not last as long as expected. The batteries installed in the PS500 depleted much earlier than expected although the battery indicator showed a sufficiently charged battery. The sequential alarms, "Battery low" followed by" Battery depleted" were given, however, the remaining time until the ventilator shut down was not foreseeable. When the battery depleted totally, the power fail alarm was generated.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2015