Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18826 |
| Brand Name: | Draeger Babylog VN500 Ventilator with optional PS500 installed |
| Model: | VN500 |
| Affected: | All units with optional PS500 installed |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Brand Name: | Draeger Infinity V500 Ventiilator with optional PS500 installed |
| Model: | V500 |
| Affected: | All units with optional PS500 installed |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Brand Name: | Draeger Evita V500 Ventilator, with optional PS500 installed |
| Model: | V500 |
| Affected: | All units with optional PS500 installed |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | Draeger has become aware of an event in which the battery capacity of the optional PS500 of an IACS Workstation Critical Care (Evita Infinity V500, Evita V300 and Babylog VN500 ventilators) did not last as long as expected. The batteries installed in the PS500 depleted much earlier than expected although the battery indicator showed a sufficiently charged battery. The sequential alarms, "Battery low" followed by" Battery depleted" were given, however, the remaining time until the ventilator shut down was not foreseeable. When the battery depleted totally, the power fail alarm was generated. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/7/2015 |