Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18824 |
| Brand Name: | Invitrogen HLA High Resolution SSP UniTray |
| Model: | 4729010, 4729110 |
| Affected: | Selected lots |
| Software version: | |
| Recalling Organisation: | Thermo Fisher Scientific New Zealand Ltd 244 Bush Road Albany AUCKLAND |
| Contact Information: | 0800 636 327 |
| Manufacturer: | Invitrogen Corporation |
| Brand Name: | Invitrogen HLA AllSet+ Gold SSP |
| Model: | 54010D |
| Affected: | Selected lots |
| Software version: | |
| Recalling Organisation: | Thermo Fisher Scientific New Zealand Ltd 244 Bush Road Albany AUCKLAND |
| Contact Information: | 0800 636 327 |
| Manufacturer: | Invitrogen Corporation |
| Issue: | The primer mix in lane 86 (A68-03C) of specific lots of these kits was expected to produce a positive reaction with DNA samples containing the A*03:20 allele. The QC Comments provided with these kits have been updated to indicate the potential for a false negative reaction with the A*03:20 and like alleles. This has the potential to provide a mistype result as an A*03:01 allele. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/7/2015 |