Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18824
Brand Name:Invitrogen HLA High Resolution SSP UniTray
Model:4729010, 4729110
Affected:Selected lots
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:Invitrogen Corporation
Brand Name:Invitrogen HLA AllSet+ Gold SSP
Model:54010D
Affected:Selected lots
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:Invitrogen Corporation
Issue:The primer mix in lane 86 (A68-03C) of specific lots of these kits was expected to produce a positive reaction with DNA samples containing the A*03:20 allele. The QC Comments provided with these kits have been updated to indicate the potential for a false negative reaction with the A*03:20 and like alleles. This has the potential to provide a mistype result as an A*03:01 allele.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/7/2015