Detail of Recall

Type of Product:Device
Medsafe Reference:18822
Brand Name:Roche Reflotron Uric Acid Test
Model:Material Number: 10745103202
Affected:All lots
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Roche Diagnostics has reduced the hematocrit value to a maximum of 48% for the Reflotron Uric Acid test when measuring patient samples from whole blood.
The internal investigation detected deviations in Uric Acid results on Reflotron system, which can be above the internal specification of 5%, if the hematocrit values exceed 48%. This may lead to erroneously low Uric Acid results in blood samples with hematocrit values higher than 48%.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/7/2015