Detail of Recall

Type of Product:Device
Medsafe Reference:18820
Brand Name:Siemens Acuson Ultrasound System, utilising the 6C1 HD and/or 18L6 HD biopsy guidelines
Model:10041461, 10441701, 10441730 utilizing the 18L6 HD and/or 6C1 HD biopsy guidelines
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:Siemens advises that the 6C1 HD and 18L6 HD transducers demonstrate a misalignment while utilising the on-screen guidelines. This issue occurs when using the CIVCO biopsy attachments in combination with the ACUSON S Family on-screen guidelines. The biopsy needle may traverse outside the on-screen guidelines. This occurs on the 6C1 HD and 18L6 HD transducers. The potential risk is the tissue of interest may not get sampled.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2015