Detail of Recall

Type of Product:Device
Medsafe Reference:18808
Brand Name:EasyCheck Pregnancy Test Cassette
Model:PMC-ECCS40
Affected:Lot 15033101
Software version:
Recalling Organisation:Phoenix MedCare Limited
Level 4 (4a), 22 Fleet Street
Eden Terrace
Auckland 1021
Contact Information:
Manufacturer:Hangzhou Clongene Biotech Co Ltd
Issue:Individual test pouches have a misprint in the results illustration under heading "Invalid" showing an"Invalid" result to be equal to a "Negative" result.
A negative result may be interpreted as an invalid result if the IFU not consulted.


Our product is supplied in boxes of 40 tests, and an IFU is included with each box. The pouch printing was a feature we have introduced on the new funded product, as an improvement to the previous tender holder (blank pouches).
Recall action type:Recall
Recall action:Product to be modified
Level of recall:Wholesale
Date Commenced:3/7/2015