Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18808 |
| Brand Name: | EasyCheck Pregnancy Test Cassette |
| Model: | PMC-ECCS40 |
| Affected: | Lot 15033101 |
| Software version: | |
| Recalling Organisation: | Phoenix MedCare Limited Level 4 (4a), 22 Fleet Street Eden Terrace Auckland 1021 |
| Contact Information: | |
| Manufacturer: | Hangzhou Clongene Biotech Co Ltd |
| Issue: | Individual test pouches have a misprint in the results illustration under heading "Invalid" showing an"Invalid" result to be equal to a "Negative" result. A negative result may be interpreted as an invalid result if the IFU not consulted. Our product is supplied in boxes of 40 tests, and an IFU is included with each box. The pouch printing was a feature we have introduced on the new funded product, as an improvement to the previous tender holder (blank pouches). |
| Recall action type: | Recall |
| Recall action: | Product to be modified |
| Level of recall: | Wholesale |
| Date Commenced: | 3/7/2015 |