Detail of Recall

Type of Product:Device
Medsafe Reference:18806
Brand Name:Siemens ACUSON S1000, ACUSON S2000, or ACUSON S3000 ultrasound systems
Model: 10441701, 10041461, 10440017, 10441730
Affected:
Software version:VC30A, VC30B, VC30C, or VC31A
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer has recently discovered a potential miscalculation when using the Doppler manual trace measurement tool in a full screen display format. The manual trace tool may incorrectly calculate the Time Average Velocity (TAV) and subsequently incorrectly calculate the Pulsatility Index (PI).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/7/2015