Type of Product:Device
Medsafe Reference:18800
Brand Name:Philips Allura Systems, Actuator Monitor Ceiling Suspension (MCS)
Model:722010, 722011, 722012, 722013, 722022, 722023, 722024 and 722015
Affected:All serial numbers
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has received a report of a failing Monitor Ceiling Suspension (MCS) Actuator on an Allura Xper system. The actuator assembly had become detached and the monitor carriage with a flexvision monitor dropped to its lowest position.
When the Monitor Ceiling Suspension falls to its lowest position there is a possibility it might collide with a patient or personnel in the room. There is also the posssibility that the monitor ceiling suspension may collide with other equipment in close proximity to the MCS.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/7/2015