Detail of Recall

Type of Product:Device
Medsafe Reference:18785
Brand Name:Biosense Webster nMARQ Circular Irrigated Catheter
Model:D132214
Affected:all lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Biosense Webster Inc
Issue:The manufacturer has received an increased number of complaints related to a low temperature measurement anomaly at electrodes of the nMARQ® Circular Irrigated Catheter.
During the same time period, has also received three reports of deaths of patients who were treated with the nMARQ® Circular Irrigated Catheter. Two of these cases were confirmed to be caused by Atrio-Oesophageal Fistula.
No direct link could be confirmed between the low temperature issue and the Atrio-Oesophageal Fistula as it was detected in only one of the two confirmed cases. Therefore, the low temperature issue may be a secondary risk factor as it may affect power titration and delivery during radio frequency ablation.As a result of these issues, Biosense Webster is recalling the nMARQ® Circular Irrigated Catheter and is conducting a full investigation.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/7/2015