Detail of Recall

Type of Product:Device
Medsafe Reference:18784
Brand Name:Siemens Acuson Ultrasound System
Model:10041461, 10440017 and 10441730
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:When repositioning the Virtual Touch IQ region of interest from the original (default) location, the lateral position of the measurement tool may not align with the lateral position of the shear velocity data. The maximum error in position between the measurement tool and shear velocity data is 3.8 mm.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/7/2015