Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18778
Brand Name:Siemens Artis zee Floorstand, Artis zee Biplane Floorstand, and Artis zee Multi Purpose System
Model:10094135, 10094141, 10094139
Affected:SW version VC14J
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:A possible position sensor fault in the swivel base axis not being detected by the system software may cause the C-arm system to move faster than normal when movement is enabled by the operator. The system will not move of its own accord, movement must still be initiated by the operator, however the system can unexpectedly exceed its usual speed, potentially resulting in a collision.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/7/2015
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /