Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18773
Brand Name:Philips HeartStart XL+ Defibrillator
Model:861290
Affected:Serial number ranges: USO1100100 to USD1101095, US11201096 to USD1203968, US11303969 to USD1309471, US11409472 to US61414022
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Philips has identifed the following XL+ software and hardware issues:

Software:
* The XL+ may fail to complete the power on sequence and continuously reboot.
* The XL+ may either fail to power up or may shut down unexpectedly.
* The XL+ may have a software version that did not reset a fail-safe monitoring component which could cause a delay of therapy or pacing interruption.
* The XL+ may fail to generate verbal prompts in AED mode.

Hardware:
* The speaker may fail.
* The battery may not seat properly causing the XL+ to shut down unexpectedly or remain powered on and not acknowledge or charge the battery.
* The XL+ exceeds the allowable radiated emissions level for Class B CISPR11.
* The XL+ ECG signal from leads could be lost and unrecoverable.
* The XL+ SpO2 signal may lose communication and cause the device to reboot.
* The XL+ battery detection system may be disrupted and cause a false low battery alarm.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/7/2015