Detail of Recall

Type of Product:Device
Medsafe Reference:18767
Brand Name:Philips SENSE Body Coil 1.5T (used with ACS NT, Intera 1.5T, and Achieve 1.5T MR Systems)
Model:4522 131 5575x, 4522 132 1985x. 4522 132 7619x, 4598 000 5187x, 4598 000 5189x, 4598 003 590511 (where x can be a number from 1 to 9)
Affected:All lots
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The Instructions for Use state how (and how not) to position the SENSE Body Coil 1.5T and its cables in relation to the patient and MR bore in order to avoid a hazardous situation. When these instructions are not followed during an examination, the positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to RF coupling to the Quadrature Body Coil (QBC)
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/7/2015