| Type of Product: | Device |
| Medsafe Reference: | 18767 |
| Brand Name: | Philips SENSE Body Coil 1.5T (used with ACS NT, Intera 1.5T, and Achieve 1.5T MR Systems) |
| Model: | 4522 131 5575x, 4522 132 1985x. 4522 132 7619x, 4598 000 5187x, 4598 000 5189x, 4598 003 590511 (where x can be a number from 1 to 9) |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | The Instructions for Use state how (and how not) to position the SENSE Body Coil 1.5T and its cables in relation to the patient and MR bore in order to avoid a hazardous situation. When these instructions are not followed during an examination, the positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to RF coupling to the Quadrature Body Coil (QBC) |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/7/2015 |