Detail of Recall

Type of Product:Device
Medsafe Reference:18760
Brand Name:Masimo uSpO2 and HPLP Oximeter Cables
Model:1075383
Affected:All lots
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics Inc
Issue:A small number of Oximetry Cables have been identified with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally, with the same Oximetry Cable, when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device.
Recall action type:Correction
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/6/2015