Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18760 |
| Brand Name: | Masimo uSpO2 and HPLP Oximeter Cables |
| Model: | 1075383 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Respironics Inc |
| Issue: | A small number of Oximetry Cables have been identified with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally, with the same Oximetry Cable, when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device. |
| Recall action type: | Correction |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/6/2015 |