Type of Product:Device
Medsafe Reference:18755
Brand Name:GE Precision and Advantx Legacy/Legacy-D Radiographic and Fluroscopic Systems
Model:
Affected:All serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:The Spot Film Device (SFD) or Intelegent Digital Device (IDD) safety mechanism installed on the system may not engage properly at the lower range of SFD/IDD travel. This mechanism is installed to hold and prevent the SFD/IDD from falling due to a counterweight cable failure. The lower range of travel is 1.1" (28mm) from the lowest point of normal SFD/IDD vertical compression. A fall of the SFD/IDD could result in an injury to a patient or operator. There have been no reported injuries as a result of this issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/6/2015