Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18755 |
| Brand Name: | GE Precision and Advantx Legacy/Legacy-D Radiographic and Fluroscopic Systems |
| Model: | |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | The Spot Film Device (SFD) or Intelegent Digital Device (IDD) safety mechanism installed on the system may not engage properly at the lower range of SFD/IDD travel. This mechanism is installed to hold and prevent the SFD/IDD from falling due to a counterweight cable failure. The lower range of travel is 1.1" (28mm) from the lowest point of normal SFD/IDD vertical compression. A fall of the SFD/IDD could result in an injury to a patient or operator. There have been no reported injuries as a result of this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/6/2015 |