Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18738 |
| Brand Name: | Cook Lunderquist Wire Guide |
| Model: | TSCMG-35-260-7-LES |
| Affected: | E3307556 |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | William Cook Europe ApS |
| Issue: | Cook Medical has received 3 reports of products with the product label indicating the wire guide is curved (the letter "C" in the prefix TSCMG- stands for curved), however actual product inside package is a straight wire guide. This observation was made prior to use of the product as the physician is able to see the wire is not curved. The Instructions For Use (IFU) state "Upon removal from package, inspect the product to ensure no damage has occurred." |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/7/2015 |