Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18735
Brand Name:Medcom Hemo-Cath ST Hemodialysis Catheters
Model:All models
Affected:All lot numbers
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Medcomp Inc
Brand Name:Medcom Hemo-Cath LT Hemodialysis Catheters
Model:All models
Affected:All lot numbers
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Medcomp Inc
Issue:The manufacturer has updated the Instructions for Use (IFU) to include an instruction for the catheter not to be clamped. Clamping the catheter can cause holes in the catheter causing leaks potentially leading to blood loss, air emboli, infection, delays in treatment, and a surgical procedure to remove and replace the device prior to continuing the dialysis treatment.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/6/2015