Detail of Recall

Type of Product:Device
Medsafe Reference:18723
Brand Name:Stryker Neptune 2 Rover Surgical fluid/smoke waste management system
Model:0702-002-000
Affected:Serial Numbers ranging from 0803118093 through 1505400223
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Instruments
Issue:It has been identified during the installation of Neptune Rovers, there is a potential for the Rover's Diverter Valve Assembly to not actuate during docking, preventing the cleaning cycle from occurring and inhibiting prefill water from being supplied to the Neptune Fresh Water System. The Diverter Valve's material combined with the heat-treating process has the potential to cause diverter valve corrosion when the Rover is allowed to sit for a period of greater than 40 days.
Only Neptune 2 Rovers out of use for greater than 40 days have the potential of exhibiting the failure. Neptune 2 Rovers that have been used within the past 40 days are not likely to experience any issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/7/2015