Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18685
Brand Name:Topcon 3D Optical Coherence Tomography 3D OCT-1 Maestro systems
Model:TOP-470109800
Affected:Serial number range: 105104 to 105701
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Topcon Corporation
Issue:In certain usage environments, when the 3D OCT-1 is used over a prolonged period of time, dust can penetrate into the optical system and adhere to the surface of the lens. This dust accumulates gradually over time and this can cause a whitish shadow artefact to occur in centre of the image and potentially create a blurry background.
If this occurs, the reduced image quality is an obvious anomaly noticed during clinical assessment which cannot be mistaken for patient pathology.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/6/2015