Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18683 |
| Brand Name: | RaySearch RayStation Radiation therapy treatment planning system |
| Model: | |
| Affected: | |
| Software version: | 4.7 |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | When certain values are used as the highest colour table level, the corresponding isodose line for display of dose and dose difference is not displayed. When certain values are used as the highest colour table level, and when using the discrete level option, values at or above the highest colour table level will be coloured as the second highest level for dose, dose difference or PET images. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/6/2015 |