Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18657 |
| Brand Name: | Siemens Somatom Emotion 6 / 16 Slice Configuration Systems |
| Model: | Catalogue Numbers: 10165888, 10165977, 10415118, 10415119 |
| Affected: | Systems manufactured before Feb-2008 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Shanghai Medical Equipment Ltd |
| Issue: | The bolts used to fix the balance weight of the data measurement system (DMS) may show signs of metal fatigue. The function of these bolts may potentially degrade over time and impact stability. It should be noted that the rupture of a single bolt does not pose a hazard, however in the unlikely situation that all bolts break over time, this could cause the DMS balance weight to drop into the gantry. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/6/2015 |