Detail of Recall

Type of Product:Device
Medsafe Reference:18657
Brand Name:Siemens Somatom Emotion 6 / 16 Slice Configuration Systems
Model:Catalogue Numbers: 10165888, 10165977, 10415118, 10415119
Affected:Systems manufactured before Feb-2008
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shanghai Medical Equipment Ltd
Issue:The bolts used to fix the balance weight of the data measurement system (DMS) may show signs of metal fatigue. The function of these bolts may potentially degrade over time and impact stability. It should be noted that the rupture of a single bolt does not pose a hazard, however in the unlikely situation that all bolts break over time, this could cause the DMS balance weight to drop into the gantry.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/6/2015