Detail of Recall

Type of Product:Device
Medsafe Reference:18653
Brand Name:Verathon GlideScope GVL and AVL Reusable Video Laryngoscope Blades
Model:Models: GVL3, GVL4, GVL5, AVL2, AVL3, AVL4, AVL5
Affected:Selected serial number ranges
Software version:
Recalling Organisation:Teleflex Medical New Zealand
OFFICE 011, 110 CARLTON GORE ROAD
Newmarket
Auckland 1023
Contact Information:027 476 4009
Manufacturer:Verathon Medical (Canada) ULC
Issue:Verathon, manufacturer of Glidescope video laryngoscopes, is revising the Operations and Maintenance Manuals (OMM) for certain Glidescope GVL and AVL Reusable Laryngoscope Blades. The new OMMs specifically state compatibility limitations of the reusable video laryngoscope validated through simulated use cycles, using certain cleaning and disinfection agents. Also, for the purpose of cleaning/disinfecting reusable blades, the following chemicals have been removed from the OMMs:
· Bleach (sodium hypochlorite)
· Isopropyl alcohol solution
· Liquid hydrogen peroxide (7.5%)
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2015