Detail of Recall

Type of Product:Device
Medsafe Reference:18632
Brand Name:Roche clinical chemistry tests - ALT, AST, CK, CK-MB, GLDH and NH3 using the NAD(H) or NADP(H) reaction
Model:Multiple catalogue numbers
Affected:All lots
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Due to information received from the field, the potential interference of Sulfasalazine and Sulfapyridine in the Alanine Amino-Transferase (ALT) tests was investigated. The interference through Sulfasalazine and Sulfapyridine was checked for all tests using NAD(H) or NADP(H). This study was done with the following strongly elevated plasma concentrations which correspond to the CLSI-Guideline EP-7-A2:
- 754 micromol/L Sulfasalazine (300 mg/L)
- 1.2 mmol/L Sulfapyridine (299 mg/L)
The interference depends on the pharmacokinetic and the clinical status of the patient (e.g. liver and kidney function, bowel resection etc.), which makes it impossible to provide exact information about serum drug concentration and half time.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/6/2015