Type of Product:Device
Medsafe Reference:18625
Brand Name:Zimmer iASSIST Tibial Alignment Guide
Model:20-8011-013-00
Affected:120265, 120793, 120794, 130135, 130539
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Zimmer CAS
Issue:Zimmer has determined that the potential exists for the spikes on the iASSIST Tibial Alignment Guide to bend or break during insertion or extraction from the tibia during use. The bending or breakage of the spikes is due to a stress riser at the junction of the spike and the main body of the instrument in combination with the presence of repeated high cyclic bending stresses from the impaction method utilized. These stresses are significantly increased when the proper steps of the surgical technique are not followed.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/6/2015