Detail of Recall

Type of Product:Device
Medsafe Reference:18614
Brand Name:Natus neoBLUE Blanket LED Phototherapy Systems
Model:
Affected:All units
Software version:
Recalling Organisation:Scanmedics Pty Ltd
Contact Information:
Manufacturer:Natus Medical Inc
Issue:The manufacturer has advised there is a potential for the neoBLUE blanket system fibre optic pads to become discolored/degraded. This may be follwed by an eventual melting of the fiber optic bundle at the connector that is inserted in the neoBLUE blanket light box.
Recall action type:Correction
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/6/2015