Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18611
Brand Name:Vitros Performance verifier 1
Model:8067324
Affected:Lot Numbers: A3823 C4084 Q2961 T3161 V3297 and X3569
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics c/- Johnson & Johnson
13A Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 800 705
Manufacturer:
Issue:The manufacturer has observed a positive bias with the first slide used from a cartridge that has been stored in the slide supply. The bias has been observed with multiple fluids across multiple GENs and coating of slides and across the 5,1.FS, 4600 amd 5600 analyzer platforms. The frist slide bias may occur when a cartridge is loaded on an analyzer and is not used for approximately 5-14 days.
OCD has revised the Range of Means for Vitros Performance Verifier1 for unconjugated biirubin (Bu) only.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/6/2015