Detail of Recall

Type of Product:Device
Medsafe Reference:18609
Brand Name:Maquet Magnus Hybrid OR Table
Model:1180.01B2
Affected:Serial numbers: 00001 to 00100
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet GmbH
Issue:Report of a collision between the C-arm of the angiography-system and the MAGNUS OR table system.
The root cause analysis has shown that the MAGNUS column provided wrong position data of the longitudinal movement. Thus the collision calculation of the angiography-system led to a wrong result.

Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/6/2015