Detail of Recall

Type of Product:Device
Medsafe Reference:18608
Brand Name:Medtronic Navigation Cannulated Taps
Model:various
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:Cannulated taps are used during spinal procedures. When cannulated taps are used in procedures without guidewires (k-wires) the cannula may become occluded. If the occlusion is not removed at point of use and is allowed to dry and harden, attempts to remove the occlusion may become more challenging. The updated IFU now states that if there is an obstruction in the cannula that cannot be removed, the device should be considered at end of life and should be discarded.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/6/2015