Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18599 |
| Brand Name: | Philips DigitalDiagnost X-Ray System |
| Model: | |
| Affected: | |
| Software version: | 4.0.x |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | The system is designed to emit an acoustic signal (beep) upon termination of an exposure. However, if the system has been powered on for more than 12 hours, the system will no longer emit this signal. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/5/2015 |