Type of Product:Device
Medsafe Reference:18599
Brand Name:Philips DigitalDiagnost X-Ray System
Model:
Affected:
Software version:4.0.x
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The system is designed to emit an acoustic signal (beep) upon termination of an exposure. However, if the system has been powered on for more than 12 hours, the system will no longer emit this signal.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/5/2015