Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18590
Brand Name:Nucletron Oncentra Brachy Planning System
Model:
Affected:
Software version:3.x (excluding v3.0) and 4.x
Recalling Organisation:Nucletron Pty Ltd
C/O Alpha Customs Services Ltd
67 Vivian Street
New Plymouth 4310
Contact Information:+61-2-9517-1300
Manufacturer:Nucletron BV
Issue:In Oncentra Brachy (Brachy Planning (BP) module and Plan Analysis (PA) module) dose may be incorrectly calculated for Regions Of Interest (ROIs) defined on a secondary image series when using specific registration methods.
The magnitude of the dose calculation error depends on the registration transformation and can vary between 0 and 100% of the correct dose value.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/5/2015