Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18590 |
| Brand Name: | Nucletron Oncentra Brachy Planning System |
| Model: | |
| Affected: | |
| Software version: | 3.x (excluding v3.0) and 4.x |
| Recalling Organisation: | Nucletron Pty Ltd C/O Alpha Customs Services Ltd 67 Vivian Street New Plymouth 4310 |
| Contact Information: | +61-2-9517-1300 |
| Manufacturer: | Nucletron BV |
| Issue: | In Oncentra Brachy (Brachy Planning (BP) module and Plan Analysis (PA) module) dose may be incorrectly calculated for Regions Of Interest (ROIs) defined on a secondary image series when using specific registration methods. The magnitude of the dose calculation error depends on the registration transformation and can vary between 0 and 100% of the correct dose value. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/5/2015 |