Detail of Recall

Type of Product:Device
Medsafe Reference:18586
Brand Name:Smiths CADD-Solis Ambulatory Infusion Pumps
Model:Model Numbers: 2111 (grey keypad), 2112 (yellow keypad), 2120 (VIP)
Affected:Serial Number 1061043 through 1067598
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical MD Inc
Issue:Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD-Solis Medication Cassette Reservoir or Administration Set "disposable" to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (engaged) position. If the user is unable to fully engage the Cassette/ Keypad Lock the keypad will be locked and the pump cannot be started in PCA mode.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/5/2015