Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18586 |
| Brand Name: | Smiths CADD-Solis Ambulatory Infusion Pumps |
| Model: | Model Numbers: 2111 (grey keypad), 2112 (yellow keypad), 2120 (VIP) |
| Affected: | Serial Number 1061043 through 1067598 |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical MD Inc |
| Issue: | Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD-Solis Medication Cassette Reservoir or Administration Set "disposable" to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (engaged) position. If the user is unable to fully engage the Cassette/ Keypad Lock the keypad will be locked and the pump cannot be started in PCA mode. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/5/2015 |