Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18574
Brand Name:Siemens Centaur Tnl Ultra
Model:SMN: 10317708 (100 test kit), 10317709 (500 test kit)
Affected:Kit Lots ending in 088, 089, 090, 091, 093 and 094 and all future lots until the issue is resolved and a follow-up communication is issued
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has observed a system-to-system bias between the TnI-Ultra™ assay on the ADVIA Centaur® CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems with the ADVIA Centaur CP generating lower results than the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT system. Siemens is actively pursuing the cause of this issue.
This observation impacts customers who use the TnI-Ultra assay on both the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems interchangeably when interpreting serial testing in patient samples. This communication does not impact customers who use TnI-Ultra results solely from either the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT system or the ADVIA Centaur CP system.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/5/2015